Description
The Votsch VCL 4010 Environmental Chamber features a 100-liter capacity chamber. The test chamber, made from stainless steel, is illuminated and includes a port on the right side for additional testing equipment. It offers a humidity range from 10% RH to 98% RH and a temperature range from -40°C to +180°C. The chamber achieves a temperature rate of change with heating at 3.5 K/min and cooling at 5.0 K/min. Factory calibration is provided for two temperature values, and the chamber includes over/under temperature protection and an air-cooled refrigeration circuit.
The maximum load capacity is 10 kg per shelf, and the chamber door is lockable. Users can control the VCL 4010 using a touchscreen control unit, and the unit has adjustable feet for stability. The Votsch operates at a sound level of 56 dB[A]. The chamber meets the EN 60519-2, 1993 standard.
Features of the Votsch VCL 4010 Environmental Chamber:
- Chamber Capacity: 100 L
- Chamber Dimensions (W x D x H) 49 x 40 x 50 cm
- Stainless Steel Test Chamber
- Illuminated Chamber
- Port on the Right Side of the Test Chamber
- For Additional Testing Equipment
- Humidity Range 10% RH to 98% RH
- Temperature Range -4040°C to +180°C
- Temperature Rate of Change
- Heating: 3.5 K/min
- Cooling: 5.0 K/min
- Factory Calibration of 2 Temperature Values
- Over/Under Temperature Protection
- Air-Cooled Refrigeration Circuit
- Maximum Load: 10 kg /shelf
- Lockable Door
- Touchscreen Control Unit
- Adjustable Feet
- Dimensions (W x D x H) 78 x 110.5 x 119 cm
- Sound 56 dB[A]
- EN 60519-2, 1993 Standard